SPCL |
SPECIAL TEST FEATURES |
FOR HYDROGEN PEROXIDE-RESULTS IN 24 HOURS WITHOUT LABORATORY TRANSFERS |
FEAT |
SPECIAL FEATURES |
BACTERIAL SPORES ON A FILTER-PAPER CARRIER PACKAGED IN A SMALL,THRMOPLASTIC CULTURE TUBE;INSIDE THE TUBE IS A SEALED-GLASS AMPOULE OF SPECIALLY-FORMULATED SOYBEAN CASEIN DIGEST CULTURE MEDIUM CONTAINING A COLOR INDICATOR-TURNS YELLOW WHEN SPORES GROW |
AJYJ |
PACKAGE MODEL NUMBER |
EZTEST H2O2 (FOR HYDROGEN PEROXIDE) |
ALCD |
USAGE DESIGN |
STERILIZATION BIOLOGICAL INDICATOR |
AFJH |
FURNISHED ITEMS |
100 UNITS;SPORE-G.STEAROTHERMOPHILUS;THERMOPLASTIC CULTURE TUBE HOUSING GLASS AMPULE AND FILTER-PAPER CARRIER |
FTLD |
FUNCTIONAL DESCRIPTION |
TEN UNITS PER CYCLE IS COMMONLY USED;EXPOSE LOAD TO STERILIZATION CYCLE;WITH ONE NON-STERILIZED UNIT MARK POSITIVE CONTROL AND THE ONE EXPOSED, CRUSH GLASS CULTURE AMPOULE TO ACTIVE;PLACE IN INCUBATOR 55-60 DEG C 24 HOURS;IF ALL SPORES ARE KILLED THE CULTURE MEDIUM WILL REMAIN RED;IF SPORES SURVIVED MEDIUM WILL TURN YELLOW (POSITIVE) |
SUPP |
SUPPLEMENTARY FEATURES |
EXPIRATION IS 24 MONTHS FROM DATE OF MANUFACTURE;READ RESULTS EVERY 12 TO 24 HOURS;ALL POSITIVE (YELLOW) SAMPLES SHOULD BE RECORDED AND DISCARDED IMMEDIATELY. ALL YELLOW SAMPLES SHOULD BE AUTOCLAVED PRIOR TO DISPOSAL |
AJCN |
PROTECTIVE STORAGE FEATURE |
CONTROLLED ROOM TEMPERATURE |
CXCY |
PART NAME ASSIGNED BY CONTROLLING AGENCY |
EZTEST H2O2 (FOR HYDROGEN PEROXIDE;SELF-CONTAINED BIOLOGICAL INDICATOR |
AQGP |
APPROXIMATE SHELF LIFE |
24.0 MONTHS |
AGAV |
END ITEM IDENTIFICATION |
TEST EQUIPMENT FOR STERILIZATION |