| CXCY |
PART NAME ASSIGNED BY CONTROLLING AGENCY |
BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
| CYFY |
ANALYSIS TEST/MEASUREMENT RANGE |
XPO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT. |
| CYFV |
METHOD OF ANALYSIS |
POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
| EXME |
EXTERIOR CONTAINER CUBIC MEASURE |
1.00 CUBIC FEET |
| ABRN |
HEIGHT |
8.400 INCHES |
| WDTH |
WIDTH |
9.100 INCHES |
| AQXY |
TEST TYPE FOR WHICH DESIGNED |
BLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT |
| ECWT |
EXTERIOR CONTAINER WEIGHT |
27.00 POUNDS |
| ELEC |
VOLTAGE IN VOLTS |
120.0 AND 230.0 |
| CBBL |
FEATURES PROVIDED |
INTEGRAL PRINTER AND DIGITAL DISPLAY SCREEN AND SELF-CALIBRATION |
| CYFZ |
TEST SAMPLE/SPECIMEN SIZE |
0.5 ML |
| CYHM |
TEST SAMPLE/SPECIMEN TYPE |
WHOLE BLOOD |
| ACDC |
CURRENT TYPE |
AC |
| FEAT |
SPECIAL FEATURES |
UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |
| ACZB |
FREQUENCY RATING |
50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
| DPTH |
DEPTH |
18.300 INCHES |